EXT_POW AlphaGUARD runs on external power
supply
-
CHARGER
Charging the battery of AlphaGUARD
in process
-
PUMP_ERR
Flow of pump module outside of valid
control range
Check tubes
connected to the
AlphaGUARD
(ensure that there is
no interruption of
the flow e.g. by
sharply bended
tubes).
If the problem
persists, contact
your local Bertin
distributor
IV_MODE Interval mode activated -
ODL_HVerr
High voltage of Geiger-Müller detector
outside of target range
Contact your local
Bertin distributor