10. Technical specifications
Model No. BM 95
Measurement method Oscillometric, non-invasive blood pressure measurement on the
upper arm.
Single-channel ECG in freely selectable front positions / ECG sig
-
nal ground (earth) referenced.
Measurement range Cuff pressure 0 – 300 mmHg,
systolic 60 – 280 mmHg,
diastolic 30 – 200 mmHg,
Pulse 30 –180 beats/minute
ECG range/sample rate
0.05 to 40Hz/256Hz
Display accuracy Blood pressure: ± 3 mmHg or 2% of the value shown
Pulse: <± 5% of the value shown
Measurement inaccuracy Max. permissible standard deviation according to clinical testing:
Systolic 8mmHg/diastolic 8mmHg
Memory 2x 60 memory spaces
Dimensions Main unit: L 128 mm x W 128 mm x H 40 mm
ECG stick: D25 mm xH 125 mm
Weight Main unit: Approximately 430 g (without batteries, with cuff)
ECG stick: Approximately 40 g
Cuff size 22 to 42 cm
Permissible operating conditions
+10 °C to +40 °C, 30-85 % relative humidity (non-condensing)
Permissible storage conditions
-20 °C to +50 °C, 10-85 % relative humidity (non-condensing)
Power supply 4x AAA batteries
Battery life For approx. 300measurements, depending on levels of blood
pressure and pump pressure
Classification
Internal supply, IPX0, no AP or APG, continuous operation
Blood pressure: Application part, type BF
ECG stick:
Application part type CF