19
Measurement distance 0.78” to 1.18” from the measuring point
Measurement range and accuracy
of forehead temperature measure-
ment
Forehead temperature measurement
93.2 °F to 108 °F (34 °C to 42.2 °C)
Measurement accuracy 96.8 °F to 102 °F: ± 0.4 °F
(36 °C to 39 °C: ± 0.2 °C)
In the remaining measurement range ± 0.5 °F (± 0.3 °C)
Clinical repeat precision 0.41 °F (0.23 °C)
Measurement range and accuracy
Object temperature measurement
Object temperature measurement
-7.6 °F to 176 °F (-22 °C to 80 °C)
Measurement accuracy ± 4% or ± 4 °F (± 2 °C)
Memory function
Body temperature measurement
Automatically stores the last 60measurements.
Acoustic signal With device activation, during measurement, when
measurement is complete (can be set).
Display LCD display
Energy saving functions Device automatically switches off after 1 minute.
Dimensions
Width x depth x height
aprox. 7.1” x 1.9” x 1.1”
(aprox. 180.3 x 47.7 x 29.1 mm)
Weight 2.65 oz (78.5 g) (without batteries)
3.43 oz (101.5 g) (with batteries)
2 x AAA (LR03) batteries Batteries last for approx. 3000 measurements
Activated functions such as acoustic signal or illumi-
nated display also reduce the battery life.
These specifications are subject to change without notice for purpose of improvement.
This device meets the requirements of the EU Directive 93/42/EC concerning
medical devices, as well as those of the Medizinproduktegesetz (German Me-
dical Devices Act), the ASTM E 1965-98 and the European Standard EN60601-
1-2, and is subject to particular precautions with regard to electromagnetic