Issue 02/2017 Art. No. 7001-0307
Operating Manual
Translation of the original operating manual
KB (E4), KB (E6)
Cooling Incubators
with compressor technology and program control
Model Model version Art. No.
KB 53 (E4) KB053-230V 9020-0199, 9120-0199
KB 53-UL (E4) KB053UL-120V 9020-0302, 9120-0302
KB 115 (E4) KB115-230V 9020-0397, 9120-0397
KB 115-UL (E4) KB115UL-120V 9020-0398, 9120-0398
KB 240 (E6) KB240-230V 9020-0202, 9120-0202
KB 240-UL (E6) KB240UL-120V 9020-0304, 9120-0304
KB 400 (E6) KB400-230V 9020-0203, 9120-0203
KB 400-UL (E6) KB400UL-120V 9020-0305, 9120-0305
KB 720 (E6) KB720-230V 9020-0204, 9120-0204
KB 720-UL (E6) KB720UL-240V 9020-0306, 9120-0306
BINDER GmbH
Address Post office box 102
78502 Tuttlingen, Germany
Tel. +49 7462 2005 0
Fax +49 7462 2005 100
Internet http://www.binder-world.com
E-mail info@binder-world.com
Service Hotline +49 7462 2005 555
Service Fax +49 7462 2005 93 555
Service E-Mail service@binder-world.com
Service Hotline USA +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS +7 495 98815 16

Table of Contents

Questions and Answers

  • V
    Victor TaylorAug 21, 2025
    Why is my Binder Accessories temperature too high with a 'Temp. range' alarm?
    • G
      gutierrezkathleenAug 21, 2025
      If you're getting a 'Temp. range' alarm and the Binder Accessories chamber is overheating, confirm the alarm. The controller might be defective, so contacting Binder service is recommended. Another possible cause is that the controller isn't adjusted correctly, in that case you should calibrate and adjust the controller.
  • S
    Steven PayneAug 23, 2025
    What to do if my Binder KB 240-UL (E6) shows "Safety controller over- temperature" alarm?
    • A
      ahenryAug 25, 2025
      If you receive a "Safety controller over- temperature" alarm on your Binder Accessories, confirm the alarm. Check the temperature set-point and the safety controller set-point. Adjust to a suitable value if needed. Also, consider if there's too much external heat load and reduce it if possible.
  • M
    Maria BoyerAug 26, 2025
    What to do if my Binder KB 240-UL (E6) Accessories shows "Safety controller over- temperature" alarm with safety controller class 3.3?
    • B
      benderjoshuaAug 26, 2025
      If your Binder Accessories is showing a "Safety controller over- temperature" alarm with safety controller class 3.3, confirm the alarm. Verify the temperature set-point and the setting of the over-temperature safety controller, adjusting to a suitable limit value if necessary. Also, check for excessive external heat load and reduce it.
  • S
    Stephen LaraAug 28, 2025
    What to do if my Binder Accessories has excess temperature and shows "Temp. range" alarm?
    • J
      James FuentesAug 28, 2025
      If your Binder Accessories displays a "Temp. range" alarm due to excess temperature, confirm the alarm. Ensure the installation site isn't too warm, as an ambient temperature above 25 °C can cause this issue. Select a cooler installation site.
  • D
    Dr. Sharon Brown PhDAug 29, 2025
    What to do if my Binder KB 240-UL (E6) temperature is too low, but the notification "Heating active" is on?
    • C
      Christina WardAug 30, 2025
      If the temperature in your Binder Accessories is too low even though the "Heating active" notification is on, it could be due to a heating defect. In this case, it's best to contact BINDER service for assistance.
  • A
    Andrew BryanSep 1, 2025
    Why is my Binder KB 240-UL (E6) program duration longer than programmed?
    • J
      John SchmittSep 1, 2025
      If the program duration on your Binder Accessories is longer than what you've set, it might be due to inappropriate tolerances programmed. For rapid transition phases, avoid programming tolerance limits to allow for maximum heating speed.
  • A
    Amanda GreenSep 3, 2025
    What to do if my Binder Accessories temperature is too low and deviates from the set-point?
    • S
      Sheila PerezSep 3, 2025
      If your Binder Accessories has a low temperature and triggers a tolerance range alarm, confirm the alarm. Make sure the chamber door is completely closed. If the issue persists, the door gasket might be defective and need replacement. You can also check the controller function, and if necessary, contact BINDER service.
  • C
    Christopher SmithSep 5, 2025
    What to do if my Binder KB 240-UL (E6) Accessories set-point temperature is not reached after specified time?
    • W
      Wendy CamachoSep 5, 2025
      If the set-point temperature on your Binder Accessories isn't reached after the specified time, make sure the chamber door is completely closed. If the problem continues, the door gasket may be defective and require replacement. You might also need to calibrate and adjust the controller.
  • S
    Shane FisherSep 7, 2025
    Why does my Binder Accessories program stop one section too early?
    • R
      riverakennethSep 7, 2025
      If your Binder Accessories program stops one section too early, it could be due to an incomplete program line. When programming with the "ramp" setting, define the end value of the desired cycle by adding an additional section with a section time of at least one minute.
  • B
    Bryce AriasSep 9, 2025
    Why does my Binder KB 240-UL (E6) Accessories have very long heating-up times?
    • J
      Jeffery LoveSep 9, 2025
      If your Binder Accessories is experiencing very long heating-up times, it might be because the chamber is fully loaded. Try loading the chamber with fewer items or allow for longer heating-up times.

Summary

Safety

General Safety Instructions for Installation and Operation

Essential safety guidelines for chamber installation and operation.

Installation of the Equipment

Electrical Connection

Detailed instructions for connecting the chamber to the power supply.

T4.12 Chamber Controller Overview

Controller Settings

USB Data Transfer

Export Data to USB

Saving chamber data and settings to a USB storage device.

Notifications and Alarms

Alarm System Overview

Summary of potential alarm conditions and their causes.

Temperature Safety Devices

Overtemperature Safety Controller (Class 3.1)

Details on the electronic over-temperature safety controller.

Maintenance, Cleaning, and Service

Troubleshooting Guide