Operating Manual
APT.lineKT (E6)
Refrigerated Incubator
with microprocessor program controller T4.12
and Peltier Refrigerating Technology
Model
Art. No.
KT 53 (E6)
9020-0250, 9120-0250
KT 53-UL (E6)
9020-0255, 9120-0255
KT 115 (E6)
9020-0209, 9120-0209
KT 115-UL (E6)
9020-0227, 9120-0227
BINDER GmbH
Address Post office box 102
D-78502 Tuttlingen
Tel. +49 7462 2005 0
Fax +49 7462 2005 100
Internet http://www.binder-world.com
E-mail info@binder-world.com
Service Hotline +49 7462 2005 555
Service Fax +49 7462 2005 93 555
Service E-Mail service@binder-world.com
Service Hotline USA +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific +852 39070500 or +852 39070503
Service Hotline Russia and CIS +7 495 98815 17

Table of Contents

Questions and Answers

  • P
    pennyhickmanAug 17, 2025
    Why is my Binder unit without function?
    • M
      Melissa SandersAug 17, 2025
      If your Binder unit isn't working, there could be several reasons. First, check the connection to the power supply. Ensure the power supply matches the unit's voltage requirements (100-120V or 200-240V). Also, inspect the unit fuse and replace it if necessary. If the fuse blows again, it is recommended to contact BINDER service.
  • D
    Daniel AvilaAug 20, 2025
    What to do if the Binder chamber doesn’t heat up?
    • A
      Alison AndersonAug 20, 2025
      If the Binder chamber isn't heating up, the safety controller might have turned it off because the limit temperature was reached, or the safety controller is set too low. Allow the unit to cool down, then press the RESET button. Check the temperature set-point and the safety controller setting. If needed, select a more suitable limit value. Note, if the safety controller is defective, contact BINDER service.
  • B
    briannelsonAug 22, 2025
    Why is the set-point temperature not reached in my Binder?
    • T
      Timothy ConleyAug 22, 2025
      If the set-point temperature isn't reached in your Binder unit, here are a few things to check: * Make sure the unit door is completely closed. * If the door gasket is defective, replace it. * The controller may need adjustment; calibrate and adjust it.
  • L
    Lance HerreraAug 26, 2025
    What to do if my Binder chamber is heating permanently?
    • T
      Thomas SalinasAug 26, 2025
      If your Binder chamber is heating constantly and exceeding the set-point, the controller might be defective, so it is recommended to contact BINDER service. It could also be that the controller needs adjustment; in this case, calibrate and adjust the controller.
  • J
    jameswrightAug 28, 2025
    What to do if overtemperature safety controller class 3.1 responds in my Binder?
    • D
      dariussmithAug 28, 2025
      If the overtemperature safety controller class 3.1 responds, it means the safety controller set-point value has been exceeded. Check the setting of the temperature set-point and the safety controller class 3.1 set-point. If appropriate, select a suitable value. Also, too much external heat load can cause this, so reduce the heat load. Note, if the controller is defective, contact BINDER service.
  • C
    Christian ThompsonAug 31, 2025
    Why does my Binder have very long heating-up times?
    • J
      Jennifer ConwayAug 31, 2025
      If your Binder unit has very long heating-up times, it could be because the chamber is fully loaded. Try loading the unit less or consider allowing longer heating-up times.
  • C
    Chris TaylorSep 2, 2025
    What to do if low refrigerating performance in Binder?
    • D
      Dr. Samantha DouglasSep 2, 2025
      If you're experiencing low or no refrigerating performance with your Binder unit, it could be due to several factors. If the ambient temperature is above 25 °C / 77°F, select a cooler place of installation. Also, too much external heat load can cause this, so reduce the heat load. If the refrigerating system is not turned on, contact BINDER service.
  • W
    William CarrSep 5, 2025
    What to do if overtemperture safety controller responds with option safety controller class 3.3 in Binder?
    • S
      Stephen GilbertSep 5, 2025
      If the overtemperature safety controller responds with option safety controller class 3.3, it means the safety controller set-point value has been exceeded. Check the setting of the temperature set- point and of the safety controller class 3.1 set-point. If appropriate, select a suitable value. Also, too much external heat load can cause this, so reduce the heat load. If the controller is defective, contact BINDER service.
  • P
    Patrick WoodSep 8, 2025
    Why does program duration longer than programmed in Binder?
    • A
      anthony50Sep 8, 2025
      If the program duration is longer than programmed in your Binder unit, it's likely due to inappropriate tolerances being programmed. For rapid transition phases, do NOT program tolerance limits to allow maximum heating speed.
  • B
    Bradley FreemanSep 12, 2025
    Why does program stops one section too early in Binder?
    • P
      perezjillianSep 12, 2025
      If the program stops one section too early in your Binder unit, it indicates that the program line is incomplete. When programming with setting “ramp”., define the end value of the desired cycle by adding an additional section with a section time of at least one minute.