Issue 11/2010 Art. No. 7001-0026
Operating Manual
APT.line™ BD (E2)
Incubators with gravity (natural) convection
APT.line™ ED (E2)
Heating ovens with gravity convection
APT.line™ FD (E2)
Heating ovens with forced convection
with R3 Controller
BINDER GmbH
Address Post office box 102
D-78502 Tuttlingen
Tel. +49 7462 2005 0
Fax +49 7462 2005 100
Internet http://www.binder-world.com
E-mail info@binder-world.com
Service Hotline +49 7462 2005 555
Service Fax +49 7462 2005 93 555
Service E-Mail service@binder-world.com
Service Hotline USA +1 866 885 9794 or +1 631 224 4340
Service Hotline Spain +34 9492 677 23
Service Hotline Asia Pacific +852 39070500 or +852 39070503
Service Hotline Russia and CIS +7 495 98815 17

Table of Contents

Questions and Answers

  • D
    Dana SmithAug 16, 2025
    What to do if my Binder APT.line BD 23 unit has no function?
    • T
      taylormoniqueAug 16, 2025
      If your Binder unit isn't functioning at all, start by checking the power supply connection. If that's okay, inspect the unit fuse and replace it if necessary. If the fuse blows again, contact BINDER service.
  • J
    Jill SnyderAug 22, 2025
    How to turn on Binder Accessories unit if only the green stand-by LED is lit?
    • J
      Jessica DeanAug 22, 2025
      If your Binder unit is only showing the green “stand-by” LED, it's likely in stand-by mode. Press and hold the ON/OFF button until the display illuminates.
  • A
    Ashley MooreAug 25, 2025
    What to do if temperature is too high and the red alarm pilot lamp is lit in Binder APT.line BD 23 Accessories?
    • P
      pricerandyAug 26, 2025
      If the temperature inside the chamber of your BD, option ED, FD Binder Accessories unit is too high and the red alarm pilot lamp of safety device is lit, check the temperature set-point and the safety device class 3.1 settings.
  • S
    Summer GordonAug 27, 2025
    What to do if Binder Accessories unit doesn't function and the red alarm pilot lamp is lit?
    • J
      Joseph CarterAug 27, 2025
      If your ED, FD Binder Accessories unit isn't working and the red alarm pilot lamp is lit, the safety device class 2 may have shut down the chamber. Allow the chamber to cool down, then press the RESET button. Check the temperature set-point and the safety device class 2 settings, and select a suitable limit value if necessary.
  • K
    Katherine YoungAug 30, 2025
    What to do if Binder Accessories chamber doesn’t heat up and the red light is off?
    • R
      riveragregoryAug 30, 2025
      If the chamber in your Binder Accessories unit isn't heating up, and the red heating control light in the display is not lit but the controller display is working, ensure the timer is correctly programmed or set to continuous operation.
  • A
    Alexander WhiteSep 2, 2025
    Why are there deviations from heating-up times in Binder Accessories?
    • R
      Robert ScottSep 2, 2025
      To address deviations from the indicated heating-up times in your Binder oven, reduce the oven load or allow for longer heating-up times.
  • F
    fletchergregorySep 4, 2025
    Why is my Binder Accessories chamber heating permanently?
    • J
      John JonesSep 5, 2025
      If the chamber in your Binder Accessories unit is heating permanently and the set-point isn't being held, calibrate and adjust the controller.
  • R
    raymondboydSep 7, 2025
    Why does the Binder Accessories controller return to Normal Display?
    • A
      Anna PerezSep 7, 2025
      If the controller returns to the Normal Display from any level, repeat entries and enter the values rapidly, because no button was hit for more than 60 sec.
  • Z
    Zachary WhitneySep 10, 2025
    Why is the set-point temperature not reached in my Binder Accessories?
    • M
      Mary DiazSep 12, 2025
      If the set-point temperature in your Binder Accessories unit isn't reached after the specified time, make sure the unit door is completely closed. Also, check if the door gasket is defective and replace it if needed. Ensure that the controller is calibrated and adjusted correctly, and verify that the power supply matches the voltage requirements (115V or 230V).
  • E
    Edward ColeSep 12, 2025
    What to do if fan doesn’t turn in Binder APT.line BD 23 Accessories?
    • L
      Lori WolfSep 12, 2025
      If the fan in your FD Binder Accessories unit isn't turning or is turning too slowly, contact BINDER service.

Summary

1. Safety

1.3 General Installation and Operation Safety

Essential safety guidelines for installation, operation, and handling of the devices.

2. Unit Description

4. Installation

4.1 Electrical Connection

Detailed instructions for connecting the unit to the power supply, including voltage and plug types.

5. Start Up

6. Controller Settings

6.1 Temperature Set-Point Entry

How to set the desired temperature, with or without a ramp function.

6.2 Time Functions

Configuration of continuous and timer operation modes for the unit.

7. Safety Devices

7.1 Temperature Safety Device Class 2

Details on the Class 2 safety thermostat for ED and FD units, including function and setting.

7.2 Temperature Safety Device Class 3.1

Details on the Class 3.1 safety thermostat for BD units, including function and adjustment.

9. Maintenance, Cleaning, and Service

11. Troubleshooting

12. Technical Description

Binder APT.line BD 23 Specifications

General IconGeneral
BrandBinder
ModelAPT.line BD 23
CategoryAccessories
LanguageEnglish