Issue 08/2012 Art. No. 7001-0115
Operating Manual
APT.lineKB (E3.1)
Refrigerated Incubators
with microprocessor program controller RD3
Model
Art. No.
KB 23 (E3.1)
9020-0112, 9120-0112
KB 23-UL (E3.1)
9020-0113, 9120-0113
KB 53 (E3.1)
9020-0114, 9120-0114
9020-0243, 9120-0243
KB 53-UL (E3.1)
9020-0115, 9120-0115
KB 53 (115 V) (E3.1)
9020-0252, 9120-0252
KB 115 (E3.1)
9020-0116, 9120-0116
9020-0242, 9120-0242
KB 115-UL (E3.1)
9020-0117, 9120-0117
KB 115 (115 V) (E3.1)
9020-0253, 9120-0253
BINDER GmbH
Address Post office box 102
D-78502 Tuttlingen
Tel. +49 7462 2005 0
Fax +49 7462 2005 100
Internet http://www.binder-world.com
E-mail info@binder-world.com
Service Hotline +49 7462 2005 555
Service Fax +49 7462 2005 93 555
Service E-Mail service@binder-world.com
Service Hotline USA +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific +852 39070500 or +852 39070503
Service Hotline Russia and CIS +7 495 98815 17

Table of Contents

Questions and Answers

  • K
    Kevin DavisAug 20, 2025
    Why is my Binder APT.line KB unit not working?
    • A
      Adam WareAug 20, 2025
      If your Binder Accessories unit isn't functioning, there could be several reasons. First, ensure it's properly connected to a power supply. Check that the voltage matches either 115V or 230V. Then, inspect the unit fuse and replace it if necessary. If the fuse blows again, contact BINDER service.
  • K
    Kathryn CherryAug 22, 2025
    What to do if set-point temperature is not reached in Binder Accessories?
    • C
      Crystal ChenAug 22, 2025
      If the set-point temperature in your Binder Accessories is not reached after the specified time, make sure the unit door is completely closed. Also, inspect the door gasket and replace it if it's defective. Finally, calibrate and adjust the controller.
  • C
    Cesar MichaelAug 25, 2025
    What to do if Binder Accessories chamber doesn’t heat up and LED (3a) “Heating active” not lit?
    • D
      Dr. Kayla MoralesAug 25, 2025
      If the chamber of your Binder Accessories unit isn't heating up and the LED (3a) “Heating active” is not lit, allow the chamber to cool down, then press the RESET button. Check the temperature set-point and the setting of the safety controller. If needed, select a suitable limit value.
  • E
    etuckerAug 26, 2025
    Why is Binder APT.line KB chamber heating permanently and set-point not maintained?
    • A
      Anna ThomasAug 26, 2025
      If the chamber in your Binder Accessories is heating permanently and the set-point is not maintained, calibrate and adjust the controller. If the problem persists, contact BINDER service.
  • B
    Bonnie RodriguezAug 27, 2025
    What to do if mechanical safety device class 3.1 responds in Binder Accessories?
    • C
      Crystal KimAug 27, 2025
      If the mechanical safety device class 3.1 responds in your Binder Accessories (with option safety device class 3.3), check the temperature set-point and the setting of safety device class 3.1. Select a suitable limit value if needed. Also, reduce the external heat load. If these steps don't resolve the issue, contact BINDER service.
  • M
    Melinda ObrienAug 30, 2025
    What to do if mechanical safety device class 3.2 responds in Binder Accessories?
    • S
      sherrymitchellAug 30, 2025
      If the mechanical safety device class 3.2 responds in your Binder Accessories (with option safety device class 3.3), check the temperature set-point and the setting of safety device class 3.2. If appropriate, select a suitable limit value. If the issue continues, contact BINDER service.
  • J
    Jeffery GriffithSep 1, 2025
    Why are there deviations from the indicated heating-up times in Binder APT.line KB Accessories?
    • S
      Sarah AlvaradoSep 1, 2025
      If you're experiencing deviations from the indicated heating-up times in your Binder Accessories, it may be due to the oven being fully loaded. Try charging the chamber with less material or consider allowing longer heating-up times.
  • E
    elizabeth78Sep 2, 2025
    What to do if Binder APT.line KB Accessories chamber doesn’t heat up but LED (3a) “Heating active” lit?
    • E
      Erin BowersSep 3, 2025
      If the chamber of your Binder Accessories unit isn't heating up, but the LED (3a) “Heating active” is lit, contact BINDER service.
  • C
    christensenchristopherSep 4, 2025
    Why is refrigerating performance low in Binder APT.line KB?
    • R
      Randy RoblesSep 5, 2025
      If you're experiencing low refrigerating performance with your Binder Accessories, it could be due to several factors. Ensure the ambient temperature is not greater than 20 °C by selecting a cooler location for installation. Reduce the heat load. If the problem persists, contact BINDER service.
  • T
    Tammy MartinezSep 6, 2025
    Why is the program duration longer than programmed in Binder APT.line KB Accessories?
    • D
      Deborah HuntSep 6, 2025
      If the program duration is longer than programmed in your Binder Accessories, it may be because inappropriate tolerances have been programmed. For rapid transition phases, do NOT program tolerance limits in order to allow maximum heating speed.

Binder APT.line KB Series Specifications

General IconGeneral
BrandBinder
ModelAPT.line KB Series
CategoryAccessories
LanguageEnglish