Issue 11/2010 Art. No. 7001-0026
Operating Manual
APT.line™ BD (E2)
Incubators with gravity (natural) convection
APT.line™ ED (E2)
Heating ovens with gravity convection
APT.line™ FD (E2)
Heating ovens with forced convection
with R3 Controller
BINDER GmbH
Address Post office box 102
D-78502 Tuttlingen
Tel. +49 7462 2005 0
Fax +49 7462 2005 100
Internet http://www.binder-world.com
E-mail info@binder-world.com
Service Hotline +49 7462 2005 555
Service Fax +49 7462 2005 93 555
Service E-Mail service@binder-world.com
Service Hotline USA +1 866 885 9794 or +1 631 224 4340
Service Hotline Spain +34 9492 677 23
Service Hotline Asia Pacific +852 39070500 or +852 39070503
Service Hotline Russia and CIS +7 495 98815 17

Table of Contents

Questions and Answers

  • R
    Rebecca ElliottAug 17, 2025
    What to do if my Binder Accessories unit has no function?
    • M
      Mark MillerAug 17, 2025
      If your Binder unit isn't functioning at all, start by checking the power supply connection. If that's okay, inspect the unit fuse and replace it if necessary. If the fuse blows again, contact BINDER service.
  • J
    James EllisAug 19, 2025
    Why is the set-point temperature not reached in my Binder APT.line FD 23 Accessories?
    • L
      Lauren SmithAug 20, 2025
      If the set-point temperature in your Binder Accessories unit isn't reached after the specified time, make sure the unit door is completely closed. Also, check if the door gasket is defective and replace it if needed. Ensure that the controller is calibrated and adjusted correctly, and verify that the power supply matches the voltage requirements (115V or 230V).
  • N
    nancy64Aug 22, 2025
    How to turn on Binder Accessories unit if only the green stand-by LED is lit?
    • J
      Jacob AdamsAug 22, 2025
      If your Binder unit is only showing the green “stand-by” LED, it's likely in stand-by mode. Press and hold the ON/OFF button until the display illuminates.
  • S
    Sheri HeathAug 26, 2025
    What to do if temperature is too high and the red alarm pilot lamp is lit in Binder APT.line FD 23 Accessories?
    • D
      David MejiaAug 26, 2025
      If the temperature inside the chamber of your BD, option ED, FD Binder Accessories unit is too high and the red alarm pilot lamp of safety device is lit, check the temperature set-point and the safety device class 3.1 settings.
  • K
    Kenneth RobertsAug 28, 2025
    What to do if Binder APT.line FD 23 Accessories unit doesn't function and the red alarm pilot lamp is lit?
    • C
      Chad AndersonAug 28, 2025
      If your ED, FD Binder Accessories unit isn't working and the red alarm pilot lamp is lit, the safety device class 2 may have shut down the chamber. Allow the chamber to cool down, then press the RESET button. Check the temperature set-point and the safety device class 2 settings, and select a suitable limit value if necessary.
  • V
    Valerie AndersonAug 30, 2025
    What to do if Binder APT.line FD 23 chamber doesn’t heat up and the red light is off?
    • D
      Danielle EdwardsAug 31, 2025
      If the chamber in your Binder Accessories unit isn't heating up, and the red heating control light in the display is not lit but the controller display is working, ensure the timer is correctly programmed or set to continuous operation.
  • C
    Craig StewartSep 2, 2025
    Why are there deviations from heating-up times in Binder APT.line FD 23 Accessories?
    • S
      Steven MayerSep 3, 2025
      To address deviations from the indicated heating-up times in your Binder oven, reduce the oven load or allow for longer heating-up times.
  • K
    Kelly BellSep 5, 2025
    Why is my Binder Accessories chamber heating permanently?
    • A
      Alexander BurnsSep 5, 2025
      If the chamber in your Binder Accessories unit is heating permanently and the set-point isn't being held, calibrate and adjust the controller.
  • M
    Michael NormanSep 8, 2025
    Why does the Binder Accessories controller return to Normal Display?
    • T
      taylorcourtneySep 8, 2025
      If the controller returns to the Normal Display from any level, repeat entries and enter the values rapidly, because no button was hit for more than 60 sec.
  • S
    Savannah BushSep 12, 2025
    What to do if fan doesn’t turn in Binder APT.line FD 23 Accessories?
    • V
      vhoffmanSep 12, 2025
      If the fan in your FD Binder Accessories unit isn't turning or is turning too slowly, contact BINDER service.

Summary

Safety and Precautions

Type Plate and General Safety

Details type plate information and general safety for installation and operation.

Intended Use

Defines the approved applications for Incubators BD and Heating Ovens ED/FD.

Installation and Connection

Electrical Connection

Details electrical connection requirements and plug types for various unit models.

Operation and Control

Advanced Controller Settings

Covers temperature set-points, ramp functions, time functions, and user level adjustments.

Safety Devices and Options

Temperature Safety Devices

Describes Class 2 and Class 3.1 temperature safety devices and their functions.

Maintenance, Disposal & Troubleshooting

Troubleshooting Guide

Lists common faults, causes, and recommended corrective actions.

Technical Specifications

Contamination Clearance Certificate

Binder APT.line FD 23 Specifications

General IconGeneral
Volume23 liters
CategoryAccessories
SeriesAPT.line
ModelFD 23
Capacity23 liters
BrandBinder
Temperature range5 °C above ambient temperature to 300 °C
Power supply230 V
Power consumption800 W
MaterialStainless steel