EasyManuals Logo

Bio-Logic SP-200 User Manual

Default Icon
122 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
Manual for VMP300-based instruments
39
Fig. 48: Loading of the flash.
5) Click on the “Browse” button to select to the flash file (EC0368.flash) of the older
version (here: V10.12). Click on “Open” when the flash file is selected.
NOTE:
if the wrong file is selected, the communication board may be damaged.
Fig. 49: Flash location in the EC-Lab folder.
6) The pathway of
the flash of the old
version (V10.12) is
now shown in the
“Firmware
Upgrade/Downgra
de” window.
Fig. 50: “Firmware Upgrade/Downgrade” window.
7) Click on the “Load”
button. A warning
message shows
up. Click on “OK”
to continue the
downgrade.
Fig. 51: Warning message.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bio-Logic SP-200 and is the answer not in the manual?

Bio-Logic SP-200 Specifications

General IconGeneral
BrandBio-Logic
ModelSP-200
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals