Plumbing the NGC Fraction Collector
Use the supplied tubing and union to plumb from the NGC System
to the dispense head on the fraction collector. Plumb the waste line
from the dispense head to the waste bottle via the union.
1
Open the dispense head cover using a small
Phillips screwdriver.
2
Connect the sample tube to the dispense head. Use the
cutout in the back of the cover to insert a fittings tightener
and tighten.
3
Attach the waste line from the dispense head waste port to
the union on the side of the NGC System.
4
Route both tubes through the top of the dispense head
cover by sliding them into the rubber grommet. Finally, close
the dispense head cover and fasten with the screw.
5
Attach the tubes to the union so the direction of flow is as
shown below.
Sample tube
To FC
inlet port
From NGC
From FC
waste port
To waste
bottle
Dispense head
waste port
Dispense head
cover
Waste line
Rubber
grommet
Cutout
Dispense head