Type | Chromatography System |
---|---|
Application | Protein purification |
Flow Rate Range | 0.001 to 100 mL/min |
UV Wavelength Range | 190 to 800 nm |
pH Range | 2 to 12 |
Sample Injection | Automated sample injection |
Temperature Control Range | 4°C to 40°C |
Software Compatibility | UNICORN |
Outlines the process of installing and plumbing the NGC Fraction Collector to your NGC System.
Unpack the fraction collector, lift from the base, and check for tubing kit and power supply.
Connect the power supply and USB cable from the FC to the NGC Chromatography System.
Open the dispense head cover using a small Phillips screwdriver for access.
Connect sample tube to dispense head and waste line to union on NGC System.
Route tubes through the cover grommet, close cover, and fasten with screw.
Place the plastic drip tray and metal rack tray onto the FC platform correctly.
Turn on the fraction collector and adjust dispense head z-height for collection vessels.
List of catalog numbers and descriptions for related Peltier rack modules and tube racks.