SPECIFICATIONS
Industry exclusive two-year warranty
MEDICAL SPECTROMETER HARDWARE
Computer:
Atomlab 950PC:
PC for Windows
®
XP with P4, 2.8 GHz processor, 256 megabyte RAM, 40 gig HD, 24x CD-RW and 3.5" floppy drive.
Equipped with: 15" Flat Color Monitor, Sound Card and Speakers, Standard Keyboard, Trackball and Hewlett-
Packard Desk Jet Printer.
Multi-Channel Analyzer
Channels: 1024
Inputs: Probe and well
Spectral Resolution: FWHM 10%
Count Rate: (Maximum) 100,000 cps
Count Rate Stability: 99%
Gross Count Rate Linearity: Within 5% up to 100,000 cps
Pulse Height Linearity: Within 2% (independent of detector)
Connectors: Signal (BNC); high voltage (MHV)
Power Supply: Regulated from 775-1225 VDC at 2 mAmps
Detector High Voltage Adjustment: Automatic H.V. adjustment for both
probe & well. Uses 10 mCi Cs-137 as the calibration source.
MEDICAL SPECTROMETER SOFTWARE:
Programs: Thyroid Uptake, Wipe Test, Bioassay, Schilling Test, Hematology, Administration/ QA, Manual MCA
Radionuclides:
Factory Programmed: Au-198, Ba-133, Co-57, Co-58, Co-60, Cr-51, Cs-137, Fe-59, Ga-67, Hg-197, I-123, I-125, I-131, In-
111, Ir-192, K-42, Na-24, Pd-103, Se-75, Sr-85, Tc-99m, TI-201, Yb-169.
User Set: Unlimited user defined isotopes, setting ROI, half life, name,
efficiency and gain.
OTHER HARDWARE:
Probe: 2" x 2" Nal (T1) integral line scintillation detector with tube base
Uptake Stand:
Dimensions: 42" l x 31" w x 62" h (106.7 x 78.7 x 157.5 cm)
Collimated Shield: Flat field collimator meeting IAEA specifications
Arm: Counterbalanced, two section arm, moves 22.5" vertically and extends 29" horizontally
from stand's vertical column.
Casters: 3" maxi-lok
Weight: 296 lb (134.3 kg)
Optional:
Well Counter:
Detector: 2" x 2" Nal (T1) integral line scintillation detector with a .75"
diameter x 1.44" deep well (1.9 x 3.7 cm)
Lead Shielding: 1" thick (2.5 cm)
Cover: .125" thick (.32 cm)
Connectors: Signal (BNC); high voltage (MHV)
Well Counter:
Lead Shielding: 2" thick (5 cm)
Cover: .125" thick (.32 cm)
Certification: ETL Listed to UL 2601 Std. and CAN CSA C22.2 No. 601.1-M90, and CE marked.