This mode is used to verify any Iodine 131, 123 or 125 concentration for staff or a patient who have direct
contact with radioactive iodine or radioactive patients. A record may then be printed out for an
individual, as a Bioassay History, or as a Bioassay Employee Summary.
NOTE: In Bioassay and Wipe Test mode, the activity shown is NET activity. This means that the background count
is subtracted from the count before the activity is calculated. For all other modes, counts shown are the actual counts
detected and are not netted for background.
NOTE: You must use the Isotope Editing screen to make isotopes available for Bioassay use, and the Isotope
Efficiencies screen to enter the Detector Efficiency for each isotope. Refer to chapter 3 for details.
7-1 BIOASSAY
7-1. The Bioassay Screen.
7. BIOASSAY