7. Click <OK> and the report will automatically print. The system returns to the Report Generation
screen from which you can print another report if you like. To close the Report Generation screen,
click on the <Close> icon.
A sample Chi-Square counting screen is shown as well as a Chi-Square printed report.
Administration Report
(See Figure 3-7.)
Following daily calibration, Chi-Square, or High Voltage, the user can print an Administration Report
showing the High Voltage Value, Calibration Factor full width half max (FWHM), most recent Chi-
Square results, the person performing the calibration and comments. This report can be printed anytime.
To Print An Administration Report
1. Click on the <Report> icon in the primary tool bar.
2. Choose <Administration Summary>. You can enter a technologist and comments. If you desire to
look at the report before printing, click on <Preview>.
3. Press <Print> to print the report. This brings up the printer screen.
4. Click on <OK> to print.
NOTE: The High Voltage Report and the Atomlab 950 Administration Report are similar. The Atomlab 950
Administration Report includes Chi-Square results, the High Voltage Report does not include the Chi-Square
results.
Isotope Editing
(See Figures 3-8, 3-9.)
This option allows the user to select and view a preset isotope, add a new isotope to the isotope list, edit
an existing user-entered isotope, and activate an isotope so that it can be used for Bioassay or Wipe
functions.
The Atomlab 950 isotope list comes with 23 preset isotopes. For these factory isotopes, Lower Level
Discriminator (LLD), Upper Level Discriminator (ULD), regulatory alarm levels and Gain are factory
preset and cannot be changed. For all user entered isotopes, LLD, ULD and Gain can be edited.
While this option is active, the currently selected isotope is always highlighted in the isotope name at the
top left of the page.
NOTE: Placing a checkmark in the <Use For> field at the bottom right of the screen allows the selected isotope to be
used for Bioassay or Wipe. If you do not check the <Use For> fields the isotope will be unavailable for these
functions. Once you have checked the appropriate function(s), add a detector efficiency as described later in this
chapter.
To Select a New Isotope
1. Click on the <Isotope> icon in the primary tool bar. The Isotope Editing screen should now be
displayed.
2. The Isotope Editing screen displays information about the currently selected isotope. Click on
<Close> to keep the currently selected isotope and return to the Main Menu.