Patient monitor Users manual
14
defibrillation
A.6.4 SpO2
BLT SpO
2
SpO
2
Measurement technic Digital SpO
2
technic
Measurement range 0% to 100%
Resolution 1%
Accuracy
70% to 100%: ±2%
0% to 69%: unspecified
Alarm range
0% to 100%
, high/low limit can be adjusted continuously.
Average time Normal:8s, slow: 16sfast: 4s
Anti-interference
ability
Anti-interference of movement
Anti-interference of electrocautery unit
Resisting low
perfusion ability
With powerful ability of resisting low perfusion, PR
amplitude can reach to 0.2% with value of SpO
2
displaying
PR modulation tone
Pitch Tone
with
SpO
2
alarm range
0% to 100%, high/low limit can be adjusted continuously.
Sensor
Pulse oximetry sensors contain LEDs that emit red light at a
wavelength of approximately 660 nm and infrared light at a
wavelength of approximately 905 nm.
The total optical output power of the sensor LEDs is less than
15 mW.
This information may be useful to clinicians, such as those
performing photodynamic therapy.
PR
Measurement range 20bpm to 250 bpm
Resolution 1 bpm
Average time 8s
Accuracy ±1% or ±1 bpm, whichever is the greater
Alarm range
0bpm300bpm, high/low limit can be adjusted continuously.
PR alarm range
0 bpm to 300 bpm, high/low limit can be adjusted