Patient monitor Users manual
16
Accuracy
±3 bpmnon-motion conditions
±5 bpmmotion conditions
Resolution 1 bpm
PI
Measurement range 0.05% to 20%
A.6.5 Temp
Standard ISO 80601-2-56
Measurement way Thermal resistance way
No. of channels 8
Measurement Range 0.0~50.0 (32~122)
Measurement Accuracy 0.0-50.0: ±0.1 (not including the probe)
Resolution 0.1
Temperature unit Centigrade (), Fahrenheit (℉)
Updated time Every 1~2 s
Self-test About every 5~10 mins
Temperature probe nominal
resistance
2522Ω@25
Temperature probe type
YSI400 series probes or other compatible probes
(accuracy ± 0.1)
Alarm Range
0.0~50.0, with upper and lower limits continuously
adjustable
Alarm Indication
Blinking display of the data and parameters, text
prompts, Three levels of alarming: sound-light
alarming, alarming with blinked data and parameters,
and that with text prompts.
Recovery time after
defibrillation
<5s
A.6.6 IBP
No. of channels 8
Sensitivity of
transducer
5uV/V/ mmHg, ±2%