This unit has received an A/A rating according to the B.H.S protocol and is “recommended for home use.” This
is the highest grading available for blood pressure monitors. Please see the B.H.S website at www.bhsoc.org.
Product using the identical measurement algorithm was tested by unaffiliated researchers using B.H.S. study
protocol. Results on file and available upon request.
10. Technical specifications
Weight: Approximately 370 gm (not including batteries)
Size: 98 (W) x 163 (L) x 106 (H) mm
Storage temperature: –20°C to 55°C/–4°F to 131°F
Humidity: 15 to 90% relative humidity maximum
Operation temperature: 10 to 40°C
Humidity 15% to 90% relative humidity maximum
Display: LCD-Display (Liquid Crystal Display)
Measuring method: Oscillometric
Pressure sensor: Capacitive
Measuring range:
SYS/DIA: 30 to 280 mmHg
Pulse: 40 to 200 beats per minute
Cuff pressure display range: 0 to 299 mmHg
Memory: 99 measurements for each user, 2 users
Measuring resolution: 1 mmHg
Accuracy: Pressure within ± 3 mmHg
Pulse ± 5% of the reading
Inflation: Automatically by pump
Deflation: Silicon passive valve
Arm Circumference: 22-32 cm (medium size)
32-42 cm (large size)
Power source: 4 dry cells (batteries), UM-3, size AA 1.5V
AM-3, size AA, 1.5V
Accessories: cuff type 7052-45 for arm circumference 22 - 32 cm
or cuff type BD050 for arm circumference 32 - 42 cm
Technical alterations reserved