411829-L, System Settings April 2012 Page 3-5
3
3.3.5 Enable the λ-All Detection Mode and TLC to Step Gradient Editor
To enable the λ-all detection mode and the TLC to Step Gradient editor, you need a Spektra license.
To purchase a Spektra license, please contact your local representative. Note that the license is not
available for Isolera Prime systems.
To install a Spektra license for the λ-all detection mode and the TLC to Step Gradient editor:
1. Save the Spektra license file in a “biotage/isolera/” directory on a USB memory device.
2. Connect the memory device to the USB port located underneath the touch screen.
3. Select the General tab in the right-hand panel.
4. Press Install License in the Licenses field.
5. If the license is valid, License OK appears in the Advanced Features text box.
6. Remove the USB memory device.
7. Restart the system:
a. Press Main Menu and then Shut Down.
b. When the message It is now safe to turn off the system appears on the screen, turn off
the system. The power switch is located underneath the touch screen.
c. Turn on the system.
3.4 Configure a Network Connection
With your system connected to the network, it is possible to:
Print purification reports to a network printer with postscript support; see “Set Up a Printer and
Auto Print of Reports” on page 3-6.
Receive a report by e-mail when a purification has been completed; see “Set Up Auto Send of
Reports” on page 3-8.
Receive an e-mail when user interaction is required during a purification, e.g. when the system
run out of empty collection vessels.
Make your reports and methods available to other users by saving them in a share folder on
your network; see “Set Up File Sharing and Auto Save of Reports” on page 3-7.
Back up your database and save the system configuration and logs in a share folder on your
network; see “Back Up and Restore the System’s Database” and “Export the System
Configuration and Logs” on page 3-12.
With the system connected to your network, it is also possible to perform the following tasks
through a standard web browser (through the Isolera Remote Viewer feature):
Create a new method. If desired, the method can be based on a previous purification found
in the Results view.
Check the status of the system and the progress of a purification.
Instruct the system to start and stop collecting fractions by clicking the Collect All button
( = fractions are collected).*
Instruct the system to switch to a new collection vessel by clicking the New Fraction button.*
Instruct the system to start and end an isocratic segment by clicking the Hold button
( = isocratic hold is enabled).*
Instruct the system to pause the purification in progress by clicking the Pause button. For
safety reasons, it is only possible to pause and not resume a run from your office.*
Instruct the system to set the current UV level to 0 (zero) AU by clicking the UV Zero
button.*
View, print, and export purification results.
* You must be logged in as the user defined in the purification run.