IEEE-488 GPIB Interface
14
IDentificatioN (IDN?)
Function
Product identifies itself
Remarks
Replies command same as U2
Remote ENable (REN)
Function
Enables remote operation.
Remarks
The unit must be addressed to listen after setting REN true.
The REMOTE indicator turns on when this command is received.
Go To Local (GTL)
Function
Returns device to local operation.
Remarks
Issuing a GTL command while the device is in Local Lockout mode does not clear the
lockout condition.
The REMOTE indicator turns off.
The LISTEN indicator remains on.
Local LOckout (LLO)
Function
Disables local operation of all devices on the bus.
Remarks
REN must be true to use LLO.
LLO is cleared by setting REN false.
InterFace Clear (IFC)
Function
Terminates all bus activity and passes control to the system controller.
Remarks
All devices are set to talker and listener idle states.
LISTEN mode is canceled, and its indicator is turned off.

Table of Contents

Other manuals for BIRD THRULINE 4027F series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BIRD THRULINE 4027F series and is the answer not in the manual?

BIRD THRULINE 4027F series Specifications

General IconGeneral
BrandBIRD
ModelTHRULINE 4027F series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals