P/N : 550-ENG-OPM-EUR-R01
0.2 Shock hazards
WARNING
Unplug the unit from its power source prior to cleaning or maintenance to prevent
personal injury or equipment damage.
S
ome chemical cleaning agents may be conductive and leave a residue that may
permit a build-up of conductive dust or dirt. D
o not allow cleaning agents to
contact electrical components and do not spray cleaning solutions onto any of
these surfaces. Personal injury or equipment damage could occur.
To ensure grounding reliability, plug the AC power cord only into a properly
grounded 3-wire hospital-grade or hospital-use outlet. Do not use extension cords.
If any doubt exists as to the grounding connection, do not operate the equipment.
Personal injury or equipment damage could occur.
Do not expose the unit to excessive moisture that would allow for liquid pooling.
Personal injury or equipment damage could occur.
Do not touch the patient and signal input/output parts simultaneously
Due to the risk of electrical shock hazard, only qualified personnel with appropriate
service documentation should service the unit.
The total power of all equipment connected to the convenience outlet strip on the
the electrical requirements shown on the rear of the stand.
Otherwise, personal injury or equipment damage could occur.
Make sure the b
uilding power source is compatible with the electrical
specifications shown on the column of the stand and on the infant warmer. Failure
to do so could result in personal injury or equipment damage.
To prevent equipment damage or accidental power disconnections, do not plug a
BT-550 power cord directly to an AC wall socket when the BT-550 is mounted on a
pedestal /stand. Always provide power to the BT-550 by using the power cord
coming directly from the pedestal /stand.