EasyManuals Logo

BK Precision 830C User Manual

BK Precision 830C
113 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #83 background imageLoading...
Page #83 background image
81
*IDN?
Description: Queries the instrument ID.
Response: <instrument model>, <firmware
version>, <serial number>
LLO
Local Lockout. This means that all front panel buttons
are disabled except for button.
GTL
Go to local.
Puts the meter into the local state,
clearing the remote state and front panel lockout.
CONFigure Subsystem
CONFigure:CAPacitance [<range>]
Description: Set capacitance range.
Parameter:
<range> : Measuring unit is Farad. Parameters
are 1000p, 10n, 100n, 1u, 10u, 100u, 1m, 10m
and 50m (will be 200m range for 830C). The
meter is set to auto range when the parameter is
omitted.
Example: CONF:CAP 100u ; Set to capacitance
function and 100μF range.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BK Precision 830C and is the answer not in the manual?

BK Precision 830C Specifications

General IconGeneral
BrandBK Precision
Model830C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals