[SOURce:]CURRent:RANGe
This command sets the current range of the electronic load module. There are two current ranges, high
range (model dependent) and low range (model dependent).
When you program a range value, the load automatically selects the range that corresponds to the
value that you program. If the value falls in a region where ranges overlap, the electronic load selects
the range with the highest resolution.
NOTE: When this command is executed, the IMMediate, TRANsient, TRIGgered, and SLEW current
settings are adjusted as follows.
• If existing settings are within new range, no adjustment is made.
• If existing settings are outside new range, the levels are set to the maximum value of the new
range.
[SOURce:]CURRent:RANGe <NRf+>
0 through max. Rated current
max. current (high range)
[SOURce:]CURRent:SLEW
This command sets the slew rate for all programmed changes in the input current level of the
electronic load. This command programs both positive and negative slew rates. MAXimum sets the
slew to the fastest possible rate. MINimum sets the slew to the slowest rate.
[SOURce:]CURRent:SLEW[:BOTH] <NRf+>
A (amps per micro second)
CURR CURR:NEG CURR:SLEW:POS