EasyManua.ls Logo

BK Precision 8600 Series - Page 48

BK Precision 8600 Series
81 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
48
[SOURce:]CURRent:SLEW:POSitive
This command sets the slew rate of the current for positive transitions. MAXimum sets the slew to the
fastest possible rate. MINimum sets the slew to the slowest rate.
Command Syntax
[SOURce:]CURRent:SLEW:POSitive <NRf+>
Parameters
See specifications
Unit
A (amps per micro second)
Examples
CURR:SLEW:POS 0.0001
Query Syntax
[SOURce:]CURRent:SLEW:POSitive?
Returned Parameters
<NR3>
Related Commands
CURR:SLEW
[SOURce:]CURRent:SLEW:NEGative
This command sets the slew rate of the current for negative transitions. MAXimum sets the slew to the
fastest possible rate. MINimum sets the slew to the slowest rate.
Command Syntax
[SOURce:]CURRent:SLEW:NEGative <NRf+>
Parameters
see specifications
Unit
A (amps per micro second)
Examples
CURR:SLEW:NEG 0.0001
Query Syntax
[SOURce:]CURRent:SLEW:NEGative?
Returned Parameters
<NR3>
Related Commands
CURR:SLEW
[SOURce:]CURRent:PROTection:STATe
This command enables or disables the overcurrent protection feature.
Command Syntax
[SOURce:]CURRent:PROTection:STATe <Bool>
Parameters
0 | 1 | OFF | ON
*RST Value
OFF
Examples
CURR:PROT:STAT 1
Query Syntax
[SOURce:]CURRent:PROTection:STATe?
Returned Parameters
<NR3>
Related Commands
CURR:PROT
[SOURce:]CURRent:PROTection