Examples
CURR:TRAN:MODE TOGG
Query Syntax
[SOURce:]CURRent:TRANsient:MODE?
Returned Parameters
<CRD>
Related Commands
CURR:TRAN:ALEV TRAN
[SOURce:]CURRent:TRANsient:ALEVel
[SOURce:]CURRent:TRANsient:BLEVel
These commands specify the transient level of the input current. The transient function switches
between level A and level B.
Command Syntax
[SOURce:]CURRent:TRANsient:ALEVel <NRf+>
[SOURce:]CURRent:TRANsient:BLEVel <NRf+>
Parameters
0 through max. Rated current
Unit
A (amps)
Examples
CURR:TRAN:ALEV 5 CURR:TRAN:BLEV 0.5
Query Syntax
[SOURce:]CURRent:TRANsient:ALEVel?
[SOURce:]CURRent:TRANsient:BLEVel
Returned Parameters
<NR3>
Related Commands
CURR:
[SOURce:]CURRent:TRANsient:AWIDth
[SOURce:]CURRent:TRANsient:BWIDth
These commands specify the transient pulse width of the input current.
Note: For Continuous, Frequency is 1/ (Awid+Bwid) and Duty is Awid / (Awid+Bwid)
For pulse, pulse width = Awid
Command Syntax
[SOURce:]CURRent:TRANsient:AWIDth <NRf+>
[SOURce:]CURRent:TRANsient:BWIDth <NRf+>
Parameters
20 to 65535 us
Unit
S (seconds)
*RST Value
500uS
Examples
CURR:TRAN:AWID 0.001 CURR:TRAN:BLEV 0.02
Query Syntax
[SOURce:]CURRent:TRANsient:AWIDth?
[SOURce:]CURRent:TRANsient:BWIDth
Returned Parameters
<NR3>
Related Commands
CURR:
[SOURce:]CURRent:HIGH