125
SPECIFICATIONS
Below are some remarks in regards to all
specifications pertaining to both 878B and 879B LCR
meters.
*Specifications are subject to change without notice.
Notes:
1. Measurement performed at the test socket.
2. Measurements performed after correct open
and short calibration.
3. DUT and test leads must be properly shielded
to guard if necessary.
4. Based after 30 minutes of warm up time and
operated at 23 °C + 5 °C, <75% R.H.
5. Q value is the reciprocal of DF.
6. Accuracies based within 10% to 100% of full
scale of range; values outside of range should
be used as reference only.
7. Based on battery powered operation.
8. --- means parallel or series measurement
mode.