94
Note:
To achieve optimum precision for all L, C, and R
measurements at either the highest or lowest
ranges, calibrate the meter before testing. See
“Calibration” section for details.
Inductance Measurement
1. Press down button for one second to turn
on the meter.
2. Press (or for model 878B) button
until L is displayed on the screen to select
inductance measurement.
3. Insert an inductor into either the input sockets or
connect alligator leads into the banana jack input
terminals and connect the clips to the component
leads as illustrated in Figure 10.
4. Press button until the desired test
frequency is displayed on screen.
5. Press (or for model 878B) button
to select between D factor, Q factor, θ angle, or

Questions and Answers: