FETCh[:SCALar]:POWer[:AC][:REAL]?
This command is used to read the output power which is in the sample cache. After sending the
command, the readings will be sent to the computer. This command does not affect the
instrument settings. This command does not trigger a measurement operation, and queries
only the latest available reading. Before reading the new reading, the command returns the old
readings.
Command syntax: FETCh:POWer?
Return parameter: <NRf>
Return parameter unit: W
MEASure[:SCALar]:POWer[:AC]:APParent?
This command is used to query the actual output apparent power.
Command syntax: MEASure[:SCALar]:POWer[:AC]:APParent?
Return parameter: <NRf>
Return parameter unit: VA
Example: MEAS:POWer:APP?
FETCh[:SCALar]:POWer[:AC]:APParent?
This command is used to read the output apparent power which is in the sample cache. After
sending the command, the readings will be sent to the computer. This command does not
affect the instrument settings. This command does not trigger a measurement operation, and
queries only the latest available reading. Before reading the new reading, the command returns
the old readings.
Command syntax: FETCh:POWer:APParent?
Return parameter: <NRf>
Return parameter unit: VA
MEASure[:SCALar]:POWer[:AC]:PFACtor?
This command is used to query the actual power factor.
Command syntax: MEASure[:SCALar]:POWer[:AC]:PFACtor?
Return parameter: <NRf>
Example: MEAS:POWer:PFAC?
FETCh[:SCALar]:POWer[:AC]:PFACtor?
This command is used to read the power factor which is in the sample cache. After sending the
command, the readings will be sent to the computer. This command does not affect the
instrument settings. This command does not trigger a measurement operation, and queries
only the latest available reading. Before reading the new reading, the command returns the old
readings.
Command syntax: FETCh:POWer:PFACtor?
Return parameter: <NRf>