cache. After sending the command, the readings will be sent to the computer. This command
does not affect the instrument settings. This command does not trigger a measurement
operation, and queries only the latest available reading. Before reading the new reading, the
command returns the old readings.
Command syntax: FETCh:CURR:PEAK:MAX?
Return parameter: <NRf>
Return parameter unit: A
2.9 LIST Commands
LIST:STATe
This command is used to set the state of list mode.
Command syntax: LIST:STATe <string>
Parameter: < DISable|ENABle >
Query syntax: LIST:STATe?
Return parameter: <string>
LIST:RECall
This command is used to recall a list file.
Command syntax: LIST:RECall <NR1>
Parameter: 0 ~ 9
Example: LIST:REC 6
Query syntax: LIST:RECall?
Return parameter: <NR1>
LIST:STEP:COUNt
This command is used to set the number of steps in a list file.
Command syntax: LIST:STEP:COUNt <NR1>
Parameter: 1 ~ 100
Example: LIST:STEP:COUN 5
Query syntax: LIST:STEP:COUNt?
Return parameter: <NR1>
LIST:REPeat
This command is used to edit the number of repeat times of the list file.
Command syntaxLIST:REPeat <NR1>
Parameter: 1 ~ 10000
Example: LIST:REP 20
Query syntax: LIST:REPeat?
Return parameter: <NR1>