MEASure[:SCALar]:FREQuency?
This command is used to query the actual output frequency.
Command syntax: MEASure[:SCALar]:FREQuency?
Parameter: None
Return parameter: <NRf>
Return parameter unit: Hz
Example: MEAS:FREQ?
FETCh[:SCALar]:FREQuency?
This command is used to read the output frequency which is in the sample cache. After sending
the command, the readings will be sent to the computer. This command does not affect the
instrument settings. This command does not trigger a measurement operation, and queries
only the latest available reading. Before reading the new reading, the command returns the old
readings.
Command syntax: FETCh:FREQuency?
Return parameter: <NRf>
Return parameter unit: Hz
MEASure[:SCALar]:CURRent[:AC]:PEAK?
This command is used to query the actual output AC current peak.
Command syntax: MEASure[:SCALar]:CURRent[:AC]:PEAK?
Return parameter: <NRf>
Return parameter unit: A
Example: MEAS:CURR:PEAK?
FETCh[:SCALar]:CURRent[:AC]:PEAK?
This command is used to read the output AC current peak which is in the sample cache. After
sending the command, the readings will be sent to the computer. This command does not
affect the instrument settings. This command does not trigger a measurement operation, and
queries only the latest available reading. Before reading the new reading, the command returns
the old readings.
Command syntax: FETCh:CURR:PEAK?
Return parameter: <NRf>
Return parameter unit: A
MEASure[:SCALar]:CURRent[:AC]:PEAK:MAXimum?
This command is used to query the actual maximum output AC current peak.
Command syntax: MEASure[:SCALar]:CURRent[:AC]:PEAK:MAXimum?
Return parameter: <NRf>
Return parameter unit: A
Example: MEAS:CURR:PEAK:MAX?
FETCh[:SCALar]:CURRent[:AC]:PEAK:MAXimum?
This command is used to read the maximum output AC current peak which is in the sample