EasyManua.ls Logo

BK Precision 9801 - MEASURE and FETCH Commands

BK Precision 9801
29 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.8 MEASURE and FETCH Commands
MEASure[:SCALar]:VOLTage[:AC]?
This command is used to query the actual output AC voltage.
Command syntax: MEASure[:SCALar]:VOLTage[:AC]?
Return parameter: <NRf>
Return parameter unit: V
Example: MEAS:VOLT?
FETCh[:SCALar]:VOLTage[:AC]?
This command is used to read the output AC voltage which is in the sample cache. After sending
the command, the readings will be sent to the computer. This command does not affect the
instrument settings. This command does not trigger a measurement operation, and queries
only the latest available reading. Before reading the new reading, the command returns old
readings.
Command syntax: FETCh:VOLTage?
Return parameter: <NRf>
Return parameter unit: V
MEASure[:SCALar]:CURRent[:AC]?
This command is used to query the actual output AC current.
Command syntax: MEASure[:SCALar]:CURRent[:AC]?
Return parameter: <NRf>
Return parameter unit: A
example: MEAS:CURR?
FETCh[:SCALar]:CURRent[:AC]?
This command is used to read the output AC current which is in the sample cache. After sending
the command, the readings will be sent to the computer. This command does not affect the
instrument settings. This command does not trigger a measurement operation, and queries
only the latest available reading. Before reading the new reading, the command returns the old
readings.
Command syntax: FETCh:CURRent?
Return parameter: <NRf>
Return parameter unit: A
MEASure[:SCALar]:POWer[:AC][:REAL]?
This command is used to query the actual output active power.
Command syntax: MEASure[:SCALar]:POWer[:AC][:REAL]?
Return parameter: <NRf>
Return parameter unit: W
Example: MEAS:POWer?

Other manuals for BK Precision 9801