EasyManuals Logo

BMC E-20C-H-O User Manual

BMC E-20C-H-O
38 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
E-20C-H-O / E-20CJ-H-O User Manual V2.1
21 / 35
13.2 Options of the Patient Menu and Corresponding
Descriptions
Option
Range
Description
Humidifier
Off, 1 5
There are five humidity levels available. As the numbering
increases, the humidity rises accordingly. Off means the
humidifier is turned off. The default setting is 2
Reslex
Off, 1 3
This feature enables the device to automatically reduce the
treatment pressure when the patient exhales, so as to make
the user more comfortable. The higher the numbering is, the
more pressure the device reduces. Off means this feature
is disabled. The default setting is Off
Ramp
Time
0 Max
Ramp
In order to increase comfort and help the patient fall asleep
easily, the pressure can increase gradually, when the Ramp
feature is enabled. The ramp time during which the initial
pressure rises to the prescribed treatment pressure can be
adjusted. As you turn the Knob to the nearest point,
the numbering increases or decreases by five minutes. The
default setting is 10 minutes. The screen displays a
real-time countdown of the remaining ramp time in minutes
Delay
On / Off
When the humidifier is on, this feature allows the airflow to
continue for about 15 minutes at a low pressure (about 2 hPa)
after you press the Knob to discontinue treatment. This
will blow off the vapor left in the humidifier to avoid any
damage to the device. When this feature is set to Off, which
means it is disabled, the airflow stops instantly after you press
the Knob . The default setting is Off
Date
2000-01-01
2099-12-31
Setting date by adjusting this option
Time
00:00
23:59
Setting time by adjusting this option
Brightness
High / Low
Setting screen brightness by adjusting this option. The
default setting is High
Mask
Type
Full Face;
Nasal;
Pillow;
Other
There are three mask types available, namely Full Face
(full-face mask), Nasal (nasal mask), and Pillow (nasal pillow
mask). The default mask type is Nasal, but the patient can
choose other suitable masks as well. When selecting masks
other than the above three types of BMC masks, the patient
can identify the masks as other
iCode
iCode,
iCode QR,
iCode QR +
iCode provides access to the patient's compliance data
during a recent time period. The iCode QR mode displays
data in sequences of characters, and the iCode QR + mode
displays data in two-dimensional codes
Use Time
0
50000 h
Use Time displays how long has the device been used by the
user. The use time can be erased
About
——
Displays related information of the device (Model, SN,
Version, ID). This is read-only and cannot be edited