E-20C-H-O / E-20CJ-H-O User Manual V2.1
5 / 35
Ramp
The ramp time ranges from 0 to 60 minutes.
Sound Pressure Level
30 dB, when the device is working at the pressure of 10 hPa.
Sound Power Level
38 dB, when the device is working at the pressure of 10 hPa.
Maximum Flow
Test Pressures (hPa)
4
8
16
20
Measured Pressure at the
Patient Connection Port (hPa)
3
7
15
19
Average Flow at the Patient
Connection Port (L/min)
80
92
91
96
SpO
2
Range: 0 100%.
The margin of error for SpO
2
between 70% and 100% is ± 3%. No strict accuracy
requirements for SpO
2
below 70%.
Pulse Rate
Range: 40 240 BPM
Margin of Error: ± 3%
Wavelengths
Red: 663 nanometers
Infrared: 890 nanometers
Maximal Optical Output Power
Less than 1.5 mw maximum average.
Tube
Length: 6 ft. (1.83 m)
The Form and the Dimensions of the Patient Connection Port
The 22 mm conical air outlet complies with ISO 5356-1.
6. Available Therapy
The device delivers the following therapy:
CPAP
Delivers Continuous Positive Airway Pressure; CPAP maintains a constant level of
pressure throughout the breathing cycle. If your health care professional has prescribed
ramp for you, you can press the Ramp Button to reduce the pressure and then
gradually increase the pressure to the therapeutic pressure setting so that you can fall asleep
more comfortably.