User Guide e-CELSIUS Performance.2
EN
25
The monitor is in power supply mode. In this case, the screen is constantly switched on
without user action.
The monitor displays data temperature screen. This screen is very important for subject
monitoring. In this case, the monitor does not automatically go into standby.
In any case, to go or to come back to standby mode, you may simply press the side button of
the monitor, represented by the following logo: In this mode, the monitor is still working
and communicates with the associated pills.
6.7.12 Low battery
The monitor has a limited autonomy depending on usage (screen on / off, number of capsules
associated, etc ...).
Therefore, it is important to always check the battery level of the monitor when using the
monitor in battery mode. The orange/green LED allows you to be notified when the battery of
the monitor is in a critical state while you need to plug it in USB ASAP.
If not, however, the monitor is switched off without being recharged on time, you simply have
to put the monitor in charge (you may have to wait a few minutes before you can use the screen
again), and reset immediately the time on the monitor.
Indeed, the radiofrequency part will restart only when the date & time of the monitor has been
reset. Warning, if an invalid time is entered, the data may be corrupted.
This action of setting the time may be done directly on the monitor or through the e-
Performance Manager software. If capsules are associated, the time setting will only be
possible from the monitor interface.
While the battery is in charge, the battery logo turns purple and always filled (indicating the
state of charge and not the battery level). Once the cable is disconnected, the logo becomes
green and shows the real percentage of the battery.
6.7.13 Overview of the alarm system
The monitor combines three categories of alarms related to the state of the system.
The conditions for triggering and warning of each are summarized in the table below.
Function
controlled
Alarm condition
Delay due to
the alarm
condition
Alarm mode
Priority
Battery of the
capsule
The remaining
autonomy of the
capsule will reach a
maximum of 500
measures
30 seconds
maximum in the
case of a real
time
communication
The message "Low"
appears in the column
Batt (Picture 20:
Detailed visualization
screen of a capsule) of
detailed visualization
screen (§ 6.7.5).
Low
Battery of the
monitor
The battery of the
monitor is low or critical
Immediate
Battery status are
indicated by the LEDs
on the front side of the
monitor
Medium
Loss of
communication
Black star if some data
are missing and orange
star if the
communication with the
pills is disrupted for
Depends on the
sampling
frequency
previously set
Star Displayed
Medium*