Page #45 background image
MRI Mode
Vercise™ DBS Remote Control 4 Directions for Use
92378769-02 41 of 1157
en
Impedances Out Of Range Screen
The impedances must be within the acceptable range before MRI Mode is enabled. If the impedances are
not within the acceptable range, the Remote Control will display an error message.
1. Press to continue.
2. The Remote Control displays a new message instructing the user to review the MRI scan risks related
to abnormal impedances. Review the “Impedance Out of Range” section in the ImageReady
TM
MRI
Guidelines before proceeding. Press
to continue.