EasyManuals Logo

Boston Scientific Vercise DB-5270 User Manual

Boston Scientific Vercise DB-5270
1162 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
Stimulator Battery Messages
Vercise™ DBS Remote Control 4 Directions for Use
92378769-02 43 of 1157
en
Stimulator Battery Messages
ERI (Elective Replacement Indicator) Screen (For Non‑Rechargeable Stimulators
Only)
This screen displays when the implanted non-rechargeable Stimulator is nearing end of service.
Stimulation is provided during the ERI period. Changes made to the stimulation will not be saved, and
stimulation will not be available soon. Batteries that have lasted one year or more without entering the ERI
period will have a minimum of four weeks between entering the ERI period and reaching end of battery
life. Patients should be advised to contact their healthcare provider to report this message screen. Press
to clear this information screen. This screen will appear whenever the Remote Control is activated.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Boston Scientific Vercise DB-5270 and is the answer not in the manual?

Boston Scientific Vercise DB-5270 Specifications

General IconGeneral
BrandBoston Scientific
ModelVercise DB-5270
CategoryMedical Equipment
LanguageEnglish

Related product manuals