11 Technical specifications
44 ARC 100 Operating Manual 900-100_IFU-V1.0_11587-S0-20141209-EN
Monopolar coagulation Moderate
Values measured using non-inductive resistors
Graph of output power P [W] as a
function of the load resistance R [] for
the "Monopolar Coagulation Moderate"
setting = 50 W
Values measured using non-inductive resistors
Graph of output power P [W] as a
function of the load resistance R [] for
the "Monopolar Coagulation Moderate"
setting = 100 W
Graph of output power P [W] as a
function of the "Monopolar Coagulation
Moderate" setting. Design load
resistance = 100
Level
U (Vp)
1
150
2
190
3
250
4
295
5
317
6
360
7
370
8
400
9
420
10
440
Table of HF output voltage U [Vp] as a
function of the "Monopolar Coagulation
Moderate" setting (no load)