11 Technical specifications
900-100_IFU-V1.0_11587-S0-20141209-EN ARC 100 Operating Manual 45
Monopolar coagulation Forced
Values measured using non-inductive resistors
Graph of output power P [W] as a
function of the load resistance R [] for
the "Monopolar Coagulation Forced"
setting = 50 W
Values measured using non-inductive resistors
Graph of output power P [W] as a
function of the load resistance R [] for
the "Monopolar Coagulation Forced"
setting = 100 W
Graph of output power P [W] as a
function of the "Monopolar Coagulation
Forced" setting. Design load resistance
= 500
Level
U (Vp)
1
720
2
1040
3
1200
4
1380
5
1500
6
1680
7
1760
8
1890
9
2210
10
2340
Table of HF output voltage U [Vp] as a
function of the "Monopolar Coagulation
Forced" setting (no load)