11 Technical specifications
46 ARC 100 Operating Manual 900-100_IFU-V1.0_11587-S0-20141209-EN
Bipolar coagulation
Values measured using non-inductive resistors
Graph of output power P [W] as a
function of the load resistance R [] for
the "Bipolar Coagulation" setting =
50 W
Values measured using non-inductive resistors
Graph of output power P [W] as a
function of the load resistance R [] for
the "Bipolar Coagulation" setting = 100
W
Graph of output power P [W] as a
function of the "Bipolar Coagulation"
setting. Design load resistance = 100
Level
U (Vp)
1
117
2
158
3
195
4
227
5
246
6
270
7
290
8
315
9
335
10
346
Table of HF output voltage U [Vp] as a
function of the "Bipolar Coagulation"
setting (no load)