14 BOWA SHE SHA smoke evacuation unit MN009-940-S0 EN
Filter Installation Instructions:
The installation of the filter into the BOWA SHE SHA Smoke Evacuation System is quick and
simple.
1. Remove the filter from the shipping box and discard any protective wrapping. Examine all filters
for damage during shipping and storage. Do not install any filter with visible signs of structural
damage.
2. Insert the filter into the filter receptacle. Be sure that the filter is seated completely against the
bottom of the filter chamber and clip is fully engaged.
WARNING: This device is not intended for evacuation of fluid. If fluid is expected to be aspirated using the
filter or the BOWA SHE SHA system, fluid collection devices must be installed with the vacuum hose
assembly. Failure to install a fluid collection device may cause filter blockage and/or electrical damage.
Filter Removal Instructions:
1. After the filter has been exhausted and requires changing, turn the Smoke Evacuation System
off and disconnect any accessory tubing attached to the filter.
2. Depress the filter tab and pull the filter from the Smoke Evacuation System and dispose of in
accordance with hospital policy. The filter may be disposed of or incinerated.
3. Clean the unit with appropriate germicide prior to re-use. Follow the indicated instructions for
maintenance and installation of a new filter.
CAUTION: Using any other filter or accessory not supplied by BOWA may cause damage to the system
and/or cause the system to be inoperable and may void the warranty.
WARNING: The filter should be changed when the Filter Life Indicator shows red flashing LED (replace).
Failure to change this filter may result in decreased efficiency and contamination of the electric motor,
vacuum pump, and sound absorbing media within the system, or non-operation of smoke evacuator.