MN009-940-S0 EN BOWA SHE SHA smoke evacuation unit 19
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The model SHE SHA is intended for use in the electromagnetic environment specified below.
The customer or user of the model SHE SHA should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment -
Guidance
Portable and mobile RF
communications equipment should be
used no closer to any part of the Model
EVL including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Radiated RF
3 V/m
d = 1.7 P 80 MHz to 800 MHz
IEC 61000-4-3
80 MHz to 2.5
GHz
3 V/m
d = 2.3 P 800 MHz to 2.5 GHz
3 Vrms
d = [3.5/V1} P
Conducted RF
IEC 61000-4-6
150 kHz to 80
MHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, should be
less than the compliance level in each
frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol: