OPA-100 INSTALLATION AND SERVICE OPERATIONS
MS0100-06-0_service_manual.doc 12/22 INSTALLATION AND SERVICE OPERATIONS
3.4 SUBSEQUENT VERIFICATION
Subsequent verification of an instrument at the same location may include the following tests.
For any subsequent verification clean up the glasses and probe.
For short-term subsequent verification, perform all tests included in the initial verification
except for the power check and the warm-up check.
For short-term subsequent verification, perform a filter calibration using the filter provided with
OPA-100.
For long-term subsequent verification, perform all tests included in the initial verification.
When the instruments have been moved to a new location (e.g. change in business address
as defined by the responsible legal authority), or have undergone repairs other than the
replacement of probe, perform all tests included in the initial verification.
3.5 ROUTINE TESTING
A routine test of the instruments should consist of at least the following.
Perform an external visual control.
Clean the glass lenses and the sample probe.
Pay attention to every screen message.