10
MMEENNUU
and
OONN//OOFFFF kkeeyyss
:
When the stimulation screen is displayed, press
I
䊳
arrow key ➝ Main Menu is displayed.
Use
䊲
to select “Setup” menu.
Press
䊳
key. “Setup” menu is displayed.
Press
䊴
key. Main menu reappears in the display.
Press ON/OFF key briefly. The stimulation screen reappears.
• Check digital control knob
In the stimulation screen turn control knob clockwise; this increases the target stimulation current;
turn knob counter clockwise to reduce target stimulation current.
• LCD display: Readability check. If necessary, increase contrast by selecting Contrast menu.
Using the digital control knob, set target stimulation current = 5.00 mA.
• Check LED display with open circuit – LED lights up red.
• Check LED display with closed circuit – LED lights up green.
• Check Sound with open and closed circuit. If necessary, increase volume by selecting Volume
menu.
Enter results in equipment logbook.
3.4 Checking the accessories
• Measure 10 kΩ ± 1 % test resistance with the ohmmeter. Enter results in equipment logbook.
• Check electrode cable as described in Section 2.2. Enter results in equipment logbook.
If available:
• Connect Stimuplex® Remote Control to front of Stimuplex® HNS 12 (3-polar plug on left). The
plug connector configuration prevents any wrong polarity connection.
• Use the remote control buttons to increase target stimulation current and reset to 0.00. Enter
results in equipment logbook.