EasyManua.ls Logo

Braun Stimuplex HNS 12 - Accessory Verification

Braun Stimuplex HNS 12
16 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10
MMEENNUU
and
OONN//OOFFFF kkeeyyss
:
When the stimulation screen is displayed, press
I
arrow key Main Menu is displayed.
Use
to select “Setup” menu.
Press
key. “Setup” menu is displayed.
Press
key. Main menu reappears in the display.
Press ON/OFF key briefly. The stimulation screen reappears.
Check digital control knob
In the stimulation screen turn control knob clockwise; this increases the target stimulation current;
turn knob counter clockwise to reduce target stimulation current.
LCD display: Readability check. If necessary, increase contrast by selecting Contrast menu.
Using the digital control knob, set target stimulation current = 5.00 mA.
Check LED display with open circuit – LED lights up red.
Check LED display with closed circuit – LED lights up green.
Check Sound with open and closed circuit. If necessary, increase volume by selecting Volume
menu.
Enter results in equipment logbook.
3.4 Checking the accessories
Measure 10 k ± 1 % test resistance with the ohmmeter. Enter results in equipment logbook.
Check electrode cable as described in Section 2.2. Enter results in equipment logbook.
If available:
Connect Stimuplex® Remote Control to front of Stimuplex® HNS 12 (3-polar plug on left). The
plug connector configuration prevents any wrong polarity connection.
Use the remote control buttons to increase target stimulation current and reset to 0.00. Enter
results in equipment logbook.