2. Maintenance and technical safety checks (with battery installed)
We recommend running the following simple checks before starting up operation for the first time
and then regularly (e.g. before daily use):
--
Visual inspection of device and accessories
--
Functional check of device
2.1 Visual inspection of Stimuplex® HNS 12 including accessories
The following components must be clean and undamaged and not show any contamination:
• Housing
• Labels
• Operating controls (instant set keys)
• Keypad
• LCD display
• Controller dial (digital control knob)
• Connecting ports
• Electrode cable (e.g. insulation or electrodes not broken)
• Test resistor
If included:
• Stimuplex® Remote Control
• Stimuplex® Pen
Never use damaged equipment or accessories!
After visual inspection is successful, carry out the function test.
2.2 Functional check of Stimuplex® HNS 12
Switch on the nerve stimulator. Target stimulation current = 0.00 mA, LED is not flashing, and no
sound is audible. Battery symbol may not be flashing (battery voltage > 7 V).
• Connect electrode cable to the front of Stimuplex® HNS 12 (to the middle 4-polar plug).
The plug connector configuration prevents wrong polarity connection. – Leave the circuit open.
• Use the mA function key to set the current range to 5.00 mA – Appears in display.