.Parameters setting summary:
Para-
meter
x/xy
Factory
Set
Setting
Final
Set
P1.xy 00-15
05
Auto
-off time: no auto-off; 01-
15
minutes auto
-
off time
P2.x 0, 1, 2
2
0=
only Hold function; 1=
only Print
f
unction; 2=Print and Hold function,
P3.xy
0,1,2,
3-50
2
0=no hold function; 1=hold lager weight
reading;
2=auto release h
old function
when weight is below 10d and
auto-
hold
new
stable weight (more than 10d);
3
-50=
unchangeable reading when the
variety is within +3~50d
P4.x
0,1,2,3,4,
5,6,7
2
0=no RS232 function; 1=output display
data when
PRINT pressed; 2=
output
gross,
tare and net weight when PRINT
pressed
; 3=continuously output
display
data
; 4=
continuously output gross, tare
and net weight;
5=output display data
one time
when scale is stable; 6=
output
gross, tare and net we
ight one time
when
scale become stable
; 7=Bio-
RS232
compatible (8 data) to NCI-SP1;
P5.x
0, 1, 2, 3,
4
3
Baud rate for RS232: 0=1200bps,
1=
2400bps, 2=4800bps, 3=9600bps
,
4=19200bps