27
P6.x 0, 1, 2
0
RS232 format:
0=8N1, 1=7O1, 2=7E1
P7.xy 00-31
13
Resolution select: 500,600,750,800,
1000,1200, 1500,2000,2400,2500, 3000,
3500, 4000,
5000
, 6000, 7000, 7500,
8000, 10000, 12000, 15000, 20000,
25000, 30000, 35000, 40000, 50000,
60000,70000,75000,80000,100000
P8.x 0,1, 2
0
Division
select: 0=1, 1=2, 2=5
P9.x
0,1,2,3,4,
5
0
Decimal point in calibration:
0= x1,
1=
x0.1,
2= x0.01; 3= x0.001; 4= x0.0001;
5= x10
P10.x 0, 1
1
Calibration unit:
0=kg, 1=lb
P11.x
0, 1, 2, 3,
4, 5, 6
6
Weighing units enable: 0=only kg;
1=
only lb; 2=only lb:oz; 3=
kg or lb;
4=
kg or lb:oz; 5=lb or lb:oz; 6=
kg, lb, or
lb:oz
P12.x
0, 1, 2, 3,
4, 5, 6, 7,
3
Power-on zero-point range:
0=
calibration zero -point +1%FS;
1=
calibration zero -point +2%FS;
2=
calibration zero-point +5%FS;
3=
calibration zero-point +10%FS;
4=
calibration zero-point +20%FS;
5=
calibration zero-point +50%FS;
6=
calibration zero-point +100%FS;
7=No limitation
P13.x 0,1….14
4
Zero range for ZERO button:
0=
Power-on zero-point ±1%FS;
1=
Power-on zero-point ±2FS;
2=
Power-on zero-point ±3%FS;
3=
Power-on zero-point ±4%FS;
4=
Power-on zero-point ±5%FS;
5=
Power-on zero-point ±10%FS;
6=
Power-on zero-point ±20%FS;
7=
No limitation
8=
Power-on zero-point +1%FS
9=
Power-on zero-point +2%FS
10=Power-on zero-point +3%FS