28
11=Power-on zero-point +4%FS
12=
Power-on zero-point +5%FS
Power-on zero-point +10%FS
Power-on zero-point +20%FS
P14.x 0, 1, 2
2
Weight signal within power-on zero
point range, Choose which data as
current power
-on zero point; 0=
; 1= calibration zero-point;
P15.x 0, 1, 2
1
Weight signal not within power-on zero
point range, Choose which data as
current power
-on zero point; 0=
; 1=
-point; 2=switch-off zero-point
;
3=continuously display “
0ˉˉˉˉ”
P16.x
4, 5, 6,7,8
8
Zero tracking range: 0=0d, no tracking;
1=
+0.25d; 2= +0.5d; 3=+1d; 4=+
1.5d;
5=+2d; 6=+3d; 7= +4d; 8=+5d
P17.x 0, 1, 2, 3
2
Data filter intensity: 0=very weak,
1=
weak, 2=middle, 3=strong
P18.x 0, 1….9
1
Check weight stability range: 0=+0.5d;
1
= ± 1d ; 2=+1.5d; 3=+2d; 4=+3d
+4d ; 6=+5d ; 7=+6d ; 8=+7d; 9=+8d
P19.x 0, 1….9
1
Overload limit range: 0=FS+0d;
1=
FS+9d; 2=101%FS; 3=
105%FS; 5=110%FS;
6=120%FS;
7=150%FS; 8=200%FS; 9=No limitation
P20.x 0,1,2
2
Backlight on-off mode:
0= Backlight is always off;
1= Backlight is always on;
2= Backlight is auto on and auto off
. I
t is
auto off after 10s when scale goes to
stable and has no key operation, and it is
auto on when scale is unstable or there
’