Model 671 User’s Manual REMOTE COMMUNICATIONS
Bristol Instruments, Inc. 39
:DELTa:METHod?
Queries the state of the :METHod function.
{ START | MAXMIN }
:CALC:DELT:METH?
:DELTa:METHod { STARt | MAXMin }
Sets the state of the :METHod function to either STARt (to get start -
current values) or MAXMin (to get maximum - minimum values).
STARt
:CALC:DELT:METH STAR
:CALC:DELT:METH MAXM
:RESet
Resets the CALCulate subsystem. This command resets the Elapsed
Time counter and sets the Min, Max and Start values to the Current
Value.
:CALC:RES
:TIMe[ :ELAPsed ]?
Queries the elapsed time since the instrument was turned on or was
reset.
A time in HH:MM:SS format.
:CALC:TIM?
:CALC:TIM:ELAP?