Model 671 User’s Manual APPENDIX F – LOGGED STATUS MESSAGES
Bristol Instruments, Inc. 58
APPENDIX F – LOGGED STATUS MESSAGES
The 671 Laser Wavelength Meter reports the instrument status at each measurement as
a hexadecimal number in the log file. The hexadecimal number corresponds to a set of
status messages that are explained in the table below. For additional information about
the status messages, please contact Bristol Instruments.
This indicates the direction of the scan that each
measurement was collected on. This bit is set for left
scans, and not set for right scans.
The instrument has detected a power measurement
that is outside of the specified range of the instrument.
The instrument has detected a power measurement
that is outside of the specified range of the instrument.
Reference Laser Not
Stable
This indicates that the internal HeNe reference laser of
the model A has not stabilized yet.
This indicates that the instrument temperature is
outside of the allowed range of 0C to 50C for the
temperature sensor, or that there is a problem with the
sensor. Measurement will continue using the
temperature at the end of the allowed range closest to
the temperature sensor’s reading.
This indicates that the instrument temperature is
outside of the allowed range of 0C to 50C for the
temperature sensor, or that there is a problem with the
sensor. Measurement will continue using the
temperature at the end of the allowed range closest to
the temperature sensor’s reading.
This indicates that the ambient pressure is outside of
the allowed range of 500 to 900 mm Hg, or that there
is a problem with the pressure sensor. Measurement
will continue using the pressure at the end of the
allowed range closest to the pressure sensor’s
reading.