Model 671 User’s Manual REMOTE COMMUNICATIONS
Bristol Instruments, Inc. 41
:POWer:OFFSet?
Queries the power offset being added to power values. The power
offset is in units of dB.
{ OFF | 1 | 2 | 3 | | 19 | 20 }
:SENS:POW:OFFS?
:POWer:OFFSet { OFF | 1 | 2 | 3 | ... | 19 | 20 }
Sets the power offset to be added to power values. The power offset is
in units of dB.
OFF
:SENS:POW:OFFS OFF
:SENS:POW:OFFS 3
STATus Subsystem
The STATus subsystem { XE “Commands:STATus” } is used to retrieve status information
from the instrument and to mask or screen what status can be retrieved.
Status registers give information about the instrument and the latest measurement. Status
enable registers allow some status information to be ignored. Status information disabled
though the status enable registers will not show up in the status registers.
STATus Subsystem Functions
:QUEStionable:CONDition?
Queries the SCPI Questionable Status Register which contains bits
that indicate that one or more measurement types are of questionable
accuracy. The bits in the register are described in the table below.
An integer which is the sum of the bit values for all bits in the register
that are set.
:STAT:QUES:COND?