Essential Instructions for Measuring Equipment, Including the
European Union (Directive 2014/32/EU MID)
Although measurement transducers are not
specifically included in the MID regulations as they
do not form a complete measuring instrument
system, in accordance with Article 1 and 4, Annex
I and Annex VII, Brodie Meter Co., LLC implements
the same stringent regulations for all products and
tests to the same standards which are used for
complete measuring instrument systems.
The complete system must contain all the
necessary components to meet the requirements
of the local regulations. These components may
include pumps, air eliminators, strainers, valves,
flow computers, etc. The unit must be sealed in
accordance with the local regulations; it is the end
user’s responsibility to ensure this happens.
Flow measuring devices are provided with two
labels which specify flow ranges. The name plate
label, which includes the factory serial number,
details the operating flow range. This is the flow
range the device will operate within without
causing damage. The custody transfer label details
the working flow range associated with a particular
weights and measures approval.
It should be noted that these may not be the same;
therefore, in trade applications, the flow ranges
specified on the custody transfer label should be
followed.
Essential Instructions for Electrical Equipment, Including the
European Union (Directive 2014/30/EU and 2014/32/EU)
This unit contains Electrostatic sensitive circuit
boards. Electrostatic safety precautions should be
taken to prevent damage.
When connecting wiring it is good practice to use
shielded cable. The shield should be connected to
earth at the read out or control systems end of the
cable; the other end of the shield should not be
connected.
This wiring practice is mandatory in order to
comply with the requirements for electromagnetic
compatibility as per the EMC directive 2014/30/EU
and MID 2014/32/EU of the council of the
European Union.
It is the end user’s responsibility to ensure that all
protective covers are in place to prevent electrical
shock and/or personnel injury.